FPQ Ferrous
Debris Monitor

Rapid screening of oil samples for ferrous wear debris using a PQ Index instrument has become an established method in most used oil laboratories. It is a brilliant wear indication and provides even further information if seen in relation to other elementary analysis methods. 

There are several sources for iron wear particles in lubricants or grease such as scratched surfaces, erosion or broken parts. Often the particles are larger than 5 microns which is a size which will not be detected by laboratory methods like ICP (Inductively Coupled Plasma). 

FPQ Ferrous Debris Monitor 3
FPQ Ferrous Debris Monitor
FPQ Ferrous Debris Monitor
FPQ Ferrous Debris Monitor

A high PQ index often indicates a serious wear problem which causes large particles to appear in the sample. The device measures all ferrous particles in a sample regardless of the size. Its innovative sensor and data processing guarantee a good stability and high sensitivity.

The instrument´s standard USB connection can be used to transfer data directly into an existing LIMS system.

The FPQ has been designed to work with standard 100 ml sample bottles. This allows to measure without transferring the sample to another specialized container. The device can also be used with standard 2 ml grease pots.

Adapters for different sample vessels are available on request.

ordering
Information

OTK-CT-11224
FPQ Ferrous Debris Monitor
Range: 0 – 15000 PQ
Repeatability: +/- 5 PQ for < 500 PQ
1% for results > 500 PQ
Power supply: 230 V DC (50/60 Hz)
Test time: 10 seconds
Dimensions: 430 x 380 x 155 mm
Weight: 7.5 kg


SAM-CT-70288
100 ml Sample Bottles (288)
Neck: 38 mm
Material: PETG


SAM-CT-70289
100 ml Sample Bottles (288)
Neck: 38 mm
Material: HDPE